2019-nCov Ag Rapid Detection Kit Immuno Chromatography
单价 $235.00 / kit对比
询价 暂无
浏览 36
发货 Chinaguangdongguangzhou预售,付款后7天内
库存 10000kit起订100kit
品牌 Longsee
Name In Vitor Detection
Per Kit 200 perps
Transport RT(4?-35?)
过期 长期有效
更新 2021-03-15 15:24
 
详细信息

Product Component

Component 25 T/kit 50 T/kit 100 T/kit 200 T/kit
1 Test Cartridge 25 pcs 50 pcs 100 pcs 200 pcs
2 One-off Swab 25 pcs 50 pcs 100 pcs 200 pcs
3 Extraction Tube and Dripper 25 pcs 50 pcs 100 pcs 200 pcs
4 Sample Extraction Buffer 15 mL 30 mL 60 mL 120 mL

Note: Components in different lot number kits are not interchangeable.

Storage Condition & Expiration Date

1. 12 months at 4-35 °C.

2. Do not freeze. The test card should be used as soon as possible within 1 hour after the aluminium foil bag is opened.

Instructions for Use

Note: Reading the IFU carefully before using this reagent, and perform testing strictly in accordance with the IFU, otherwise reliable results cannot be guaranteed.

1. Before testing

It is recommended to unpack the test card after equilibrating to room temperature and use it as soon as possible within the validity period to prevent the test card from humidity before testing.

2. Extract

1)Transfer 400 μL (about 10 drops) sample extraction buffer to the extraction tube vertically.

2)Insert the swab which has collected secretions into the sample extraction buffer,rotate about 10 times and stay still at room temperature for 5 minutes.

3) Squeeze the swab tip to keep the liquid in the tube as much as possible and take out the one-off swab.

4) Cover the dripper.

3. Loading

Tear off the aluminum foil packaging bag, take out the test card, lay it flat on a horizontal table, drop 3~4 drops (about 100 μL) of the sample vertically into the sample hole, and start timing.

4. Detect

Observe the result in 15 minutes after adding the sample, and the test result will be invalid after 20 minutes.

Guangdong Longsee

Interpretation of the Result

1. Invalid: When there is no red line in the quality control area (C), the test is invalid. It is recommended to retest with a new test card, paying particular attention to whether the sample volume is sufficient.

2. Positive: A red line appears in the detection area (T) and the quality control area (C) .

3. Negative: A red line only appears in the quality control area (C).

Guangdong Longsee